A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information…
High CVSS: 8.4

CVE-2025-14096

A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system.

Other related CVE's are CVE-2025-14095 & CVE-2025-14097.



Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.





Required Configuration for Exposure:


Attacker requires physical access to the analyzer.



Temporary work Around:
Only authorized people can physically access the analyzer.

Permanent solution:
Local Radiometer representatives will contact all affected customers to discuss a permanent solution.

Exploit Status:


Researchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.